Verify your autoclave’s sterilization performance using bacterial indicator strips containing Bacillus stearothermophilus spores.
See more product details
Options:
(Select option to see volume pricing availability)
Product Details
Ensure your autoclave is operating properly with these strips for Autoclave Verification. Each strip contains Bacillus stearothermophilus spores—thermophilic, highly resistant organisms capable of growing at 65 °C and above—individually sealed in protective envelopes. During verification, place the strips in the autoclave where steam exposure occurs. After the sterilization cycle, incubate the strip in nutrient broth at 55–60 °C for at least 48 hours. The absence of microbial growth confirms effective sterilization. A reliable, easy-to-use biological indicator for routine performance checks in science labs.